FID或FPD无法点火。
请对以下项目进行安全检修:
Press (HOME) - [Detector] button on GC instrument monitor (HOME) and display the folloeing screen.
Make sure that the detector gas is properly supplied during the ignition operation.
If multiple detectors are registered in the line, press the tab to switch the detector display, or press [LINE] at the top right of the screen to switch the line.

* The upper row is the measured value, and the lower row is the set value.
Check the standard detector gas flow rate in the following table.

Set the detector temperature to 200 °C or higher (since ignition may be difficult at low detector temperature).
Press (HOME) - [Detector] button on GC instrument monitor (HOME) and display the folloeing screen.
If multiple detectors are registered in the line, press the tab to switch the detector display, or press [LINE] at the top right of the screen to switch the line.

The column is connected and the column insertion length is appropriate.
Capillary column
Packed column
Inspect the FID jet according to the following manual.
Check or replace the FID or FPD igniter filament.
Turn off the main power of the instrument, stop gas supply,
and contact your Shimadzu sales/service representative.